USB650-UV-VIS Spectrophotometer for Education

The Red Tide Spectrometer is a low-cost, small- footprint lab spectrometer that's ideal as a general-purpose instrument for budget-conscious teaching and research labs. The Red Tide has a wavelength range of 200-850 nm, and utilizes a detector with 650 active pixels; that’s 650 data points in one full spectrum, or one data point per nanometer. Configured with a 25 µm entrance slit, the Red Tide offers ~2.0 nm optical resolution (FWHM).

Features

  • Fully integrated, preconfigured system covers the 200-850 nm range at 2.0 nm optical resolution (FWHM)

  • See changes as small as 0.1 Absorbance Units

  • Small-footprint system ideal for the laboratory and the classroom

  • Direct-attach deuterium tungsten light source and sample holder

  • USB 2.0 interface for fast, hassle-free spectrometer-to-PC connection

Direct-attach Light Source

The USB650-UV-VIS features the USB-ISS-UV-VIS combination sample holder and deuterium tungsten light source (200-1100 nm). This sampling system holds standard 1-cm cuvettes for easy-to-perform absorbance measurements. The USB650-UV-VIS allows you to control both the intensity of the tungsten bulb and the shutter via software. The USB650-UV-VIS requires an external power supply (included).

Modular Design

Ocean Optics offers complete line of fiber optic sampling accessories and teaching tools to further enhance your USB650-UV-VIS system. Most accessories have SMA 905 connectors for application flexibility; changing the sampling system is as easy as unscrewing a connector and adding new components or accessories.

For even more flexibility, consider the CHEM4-UV-FIBER. An optical fiber based system, the CHEM4-UV-FIBER lets you quickly and easily couple fiber optic accessories, such as reflection probes or transmission dip probes, to the system's light source.

Plug-and-Play Operation

The USB650-UV-VIS interfaces to a PC, PLC or other embedded controllers via a USB 2.0 port. Data unique to each spectrometer is programmed into a memory chip on the USB650-UV-VIS; our spectrometer operating software reads these values for easy setup and hot swapping among PCs.

Specifications

Physical

Spectrometer dimensions: 89.1 mm x 63.3 mm x 34.4 mm
Spectrometer weight: 190 grams
Light source dimensions: 198 x 105.1 x 40.6
Light source weight: 200 g
Detector Specifications

Detector:

Sony ILX-511 Linear CCD Detector

Pixels:

650 pixels

Pixel size:

14 μm x 200 μm

Pixel well depth:

62,500 electrons

Signal-to-noise ratio: 250:1 (at full signal)
A/D resolution: 12 bit
Corrected linearity: >99.8%
Sensitivity: 75 counts at 400 NM; 41 photons/count at 600 nm
Optical Bench
Design: f/4, Asymmetrical crossed Czerny-Turner
Focal length: 42 mm input; 68 mm output
Entrance aperture: 25 µm wide slit
Grating: Grating #1
OFLV filter: OFLV-200-850
UV enhanced window: Yes, UV4
Fiber optic connector: SMA 905 to 0.22 numerical aperture single-strand optical fiber
Spectroscopic
Wavelength range: 210 to 880 nm
(The actual range typically exceeds what's listed here. The range listed is where you will see changes as small as 0.01 absorbance units.)
Optical resolution: 2.0 nm FWHM
Signal-to-noise ratio: 250:1 (at full signal)
A/D resolution: 12 bit
Dark noise: 3.2 RMS
Integration time: 3ms to 65 seconds
Dynamic range: 2 x 108 (system), 1300:1 for a single acquisition
Stray light: <0.05% at 600 nm; 0.10% at 435 nm
Light Source/Sample Holder
Power consumption: 0.6 A @ 5 VDC; external power supply (included)
Pathlength: 1 cm
Cuvette shape: Square
Light source: Deuterium tungsten
"Z" dimension: 15 mm
Bulb life (hours): 800 deuterium; 2,000 tungsten
Time to stabilized output: ~30 minutes
Electronics
Power consumption: 90 mA @ 5 VDC
Data transfer speed: Full scans to memory every 13 ms with USB 2.0 port, 18 ms with USB 1.1 port
Inputs/Outputs: None
Analog channels: No
Auto nulling: Yes
Breakout box compatible: Yes. Requires Adapter.
Trigger modes: 5 modes
Strobe functions: Yes
Connector: 10-pin connector
Computer
Operating systems: Windows 98/Me/2000/XP, Mac OS X and Linux with USB port
Computer interfaces: USB 2.0 @ 480 Mbps (USB1.1 compatible)
Peripheral interfaces: SPI (3-wire); I2C inter-integrated circuit
Pricing
Retail: $2887

 

 


PRODUCTS | TECHNICAL | APPLICATIONS | CORPORATE | DISTRIBUTORS | PRIVACY POLICY
CONTACT US
| CATALOG | PRICE LIST | ORDER INFO


.
©2007 Ocean Optics, Inc. All rights reserved.  Terms of Use and Privacy Statement
+1 727.733.2447 • +1 Fax 727.733.3962 •

 

 

 

 

Contact an Ocean Optics Applications Scientist

 

Convenient Wavelength Calibration!
Click here for Details

 

CHEMUSB4-VIS-NIR

 

CHEM4-UV-FIBER

 

CHEM4-VIS-FIBER

 

CHEM4 Systems for Education